XD7-1.05/35

Mainly for medical breast diagnosis and soft X - ray application
Product serial number
XD7-1.05/35
Product description
Parameters
Outline Drawing
Characteristics
Reserve & Transport Condition
Quality Promises After Selling
Nominal X-ray tube voltage(kV) 35
Filament characteristic(V) Ifmax=4A,Uf=3.8±0.4
Target materials Mo
Target Angle 19°
The inherent filtration(mmAl) ≤1.0
Weight(≤kg) 0.7
Focal nominal value 1
Anode nominal input power(0.1s)(kW) 0.7(small) 1.05(big)
Note1:The frequency of filament heating power supply is 0~25kHz
Note2:The filament should work normally during the warranty period of the pipe

 

We could not find any corresponding parameters, please add them to the properties table

XD7-1.05-35(5)

杭州万东电子有限公司

 

 

Load characteristic curve

杭州万东电子有限公司

FIG. 2 Load characteristic curve

Cathode emission characteristic curve

杭州万东电子有限公司

FIG. 3 Cathode emission characteristic curve

 

 

Filament characteristic curve

杭州万东电子有限公司

FIG. 4 Filament characteristic curve

Anodic heating and cooling characteristic curve

                                                                                             杭州万东电子有限公司   

FIG. 5 Anodic heating and cooling characteristics

 

 

1、 The installation or operation of the pipe must be carried out by professionals, and care must be taken during transportation and installation to avoid damaging the pipe.
2、 The compressive strength of oil shall be not less than 35kV/2.5mm as insulation and cooling medium.
3、 The pipe shell shall be cleaned and dried before installation.
4、 After the first use or disuse for more than four weeks, the pipes should be trained as follows before being put into use.From half of the maximum working tube voltage, keep the tube current value under perspective, and rise to the maximum working tube voltage at a speed no more than 5kV/min, stable for 3min.If the tube current is unstable or discharge occurs, the tube voltage should be lowered, and the tube voltage should be increased after stabilization.
5、 Before shutdown, should reduce the tube voltage, then shutdown.

 

Storage and transportation conditions

1. Generally, the packaged X-ray tube should be kept in a well-ventilated room with a temperature no less than 5℃, relative humidity no more than 80%, and no corrosive gas.

2. Transportation requirements shall be in accordance with the order contract.

 
 

Under the condition that the packaged X-ray tubes comply with the storage, transportation and use rules, our company will replace the products for customers free of charge within 12 months from the date of delivery due to manufacturing quality problems.