our PRODUCT

All categories
Number of views:
1000

XD3-3.5/100

It is mainly used for medical diagnostic X-ray machine
Product serial number
XD3-3.5/100
Product description
Parameters
Outline Drawing
Characteristics
Reserve & Transport Condition
Quality Promises After Selling
 
Nominal X-ray tube voltage(kV) 100
Filament characteristic(V) Ifmax=4.5A,Uf=7±0.8
Target materials W
Target Angle 19°
The inherent filtration(mmAl) ≤1.0
Weight(≤kg) ≤0.6
Focal nominal value 2.6
Anode nominal input power(0.1s)(kW) 3.5
Note1:The frequency of filament heating power supply is 0~25kHz
Note2:The filament should work normally during the warranty period of the pipe

 

We could not find any corresponding parameters, please add them to the properties table

XD2-1.4/85The contour diagram is shown in Figure 1 below

 

Figure 1: Outline of XD2-1.4/85

The load characteristic curve is shown in Figure 2.

杭州万东电子有限公司

Figure 2: Load characteristic curve

The emission characteristic curve of filament is shown in Figure 3.

 

 

杭州万东电子有限公司

Figure 3: Filament emission characteristic curve

 

 

 

 

 

1、 The installation or operation of the pipe must be carried out by professionals, and care must be taken during transportation and installation to avoid damaging the pipe.
2、 The compressive strength of oil shall be not less than 35kV/2.5mm as insulation and cooling medium.
3、 The pipe shell shall be cleaned and dried before installation.
4、 After the first use or disuse for more than four weeks, the pipes should be trained as follows before being put into use.From half of the maximum working tube voltage, keep the tube current value under perspective, and rise to the maximum working tube voltage at a speed no more than 5kV/min, stable for 3min.If the tube current is unstable or discharge occurs, the tube voltage should be lowered, and the tube voltage should be increased after stabilization.
5、 Before shutdown, should reduce the tube voltage, then shutdown.

Storage and transportation conditions

1. Generally, the packaged X-ray tube should be kept in a well-ventilated room with a temperature no less than 5℃, relative humidity no more than 80%, and no corrosive gas.

 

2. Transportation requirements shall be in accordance with the order contract.

 

Under the condition that the packaged X-ray tubes comply with the storage, transportation and use rules, our company will replace the products for customers free of charge within 12 months from the date of delivery due to manufacturing quality problems.

logo

这是描述信息

Copyright:Hangzhou Wandong Electron Co., Ltd 浙ICP备05030326号-1  power by:300.cn