HXD51-20、40/125

It is mainly used for medical diagnosis and matching with single-phase or three-phase full-wave rectifying high-voltage generator of central ground
Product serial number
HXD51-20、40/125
Product description
Parameters
Outline Drawing
Characteristics
Reserve & Transport Condition
Quality Promises After Selling
Maximum operating tube voltage(kV) 125
Filament characteristic(V) (Small filament)Ifmax=5.5A,Uf=7~9V
  (The filament)Ifmax=5.5A,Uf=10~13V
Target materials W
Target Angle 17.5°
Diameter(mm) 70
Speed(r/min) 2700(50Hz)
The inherent filtration(mmAl) 0.8~1.2
Weight(≤kg) 18
Focal nominal value 1.0/2.0
Maximum power(0.1s)(kw) 20/40
Continuous maximum rating(W) 410
Component radiation field 354mm*354mm
Component heat capacity 900kJ(1250kHU)
Insulation resistance between low voltage terminal and housing 2MΩ
Stator parameters Power supply frequency (HZ) 50(start)
  50(working)
Voltage(V) 160(start)
  40(working)
Maximum startup time(S) 1.2
Split phase capacitor (μF) 40(start)
  40(working)

 

We could not find any corresponding parameters, please add them to the properties table

杭州万东电子有限公司

 

 

杭州万东电子有限公司

杭州万东电子有限公司

 

 

 

 

1、 The installation or operation of the pipe must be carried out by professionals, and care must be taken during transportation and installation to avoid damaging the pipe.
2、 The compressive strength of oil shall be not less than 35kV/2.5mm as insulation and cooling medium.
3、 The pipe shell shall be cleaned and dried before installation.
4、 After the first use or disuse for more than four weeks, the pipes should be trained as follows before being put into use.From half of the maximum working tube voltage, keep the tube current value under perspective, and rise to the maximum working tube voltage at a speed no more than 5kV/min, stable for 3min.If the tube current is unstable or discharge occurs, the tube voltage should be lowered, and the tube voltage should be increased after stabilization.
5、 Before shutdown, should reduce the tube voltage, then shutdown.

Storage and transportation conditions

1. Generally, the packaged X-ray tube should be kept in a well-ventilated room with a temperature no less than 5℃, relative humidity no more than 80%, and no corrosive gas.

2. Transportation requirements shall be in accordance with the order contract.

 

Under the condition that the packaged X-ray tubes comply with the storage, transportation and use rules, our company will replace the products for customers free of charge within 12 months from the date of delivery due to manufacturing quality problems.